Marzhauser CORVUS controller
|
Summary: |
Interfaces with Marzhauser ITK Corvus Controllers and associated hardware |
|
Author: |
Johan Henriksson |
|
License: |
LGPL |
|
Platforms: |
All (uses serial interface) |
|
Devices: |
Controller, XY Stage and Z Axis |
Serial settings:
| AnswerTimeout | 500.0000 |
| DataBits | 8 |
| DelayBetweenCharsMs | 0.0000 |
| Handshaking | Off |
| StopBits | 1 |
BaudRate depends on the DIP switches:
| DIP1 | DIP2 | BaudRate |
|---|---|---|
| down | down | 9600 |
| down | up | 19200 |
| up | down | 57600 |
| up | up | 115200 |
It has been reported that an XY stage error (“Is the device is connected to the port”) can be prevented by setting the controller (and Micro-Manager) to 115200 baud and setting DelayBetweenChars to 25 ms. –Mark Tsuchida 15:35, 9 January 2015 (PST)
See also: Marzhauser (TANGO) device adapter