Edit page History How do I edit this website?

Corvus

Marzhauser CORVUS controller

Summary:

Interfaces with Marzhauser ITK Corvus Controllers and associated hardware

Author:

Johan Henriksson

License:

LGPL

Platforms:

All (uses serial interface)

Devices:

Controller, XY Stage and Z Axis

Serial settings:

   
AnswerTimeout 500.0000
DataBits 8
DelayBetweenCharsMs 0.0000
Handshaking Off
StopBits 1
   

BaudRate depends on the DIP switches:

DIP1 DIP2 BaudRate
down down 9600
down up 19200
up down 57600
up up 115200
     

It has been reported that an XY stage error (“Is the device is connected to the port”) can be prevented by setting the controller (and Micro-Manager) to 115200 baud and setting DelayBetweenChars to 25 ms. –Mark Tsuchida 15:35, 9 January 2015 (PST)