top bar

Yokogawa CSUX

(Redirected from CSUX)

Yokogawa CSUX adapter

Summary:Controls Yokogawa CSUX (all models)
Author:Nico Stuurman
License:LGPL
Platforms:Windows, Mac OS X and linux (serial port)
Devices:Yokogawa CSUX Hub, Filter Wheel, Dichroic, Shutter, DriveSpeed
Micro-Manager version:1.2.11
Serial port settings:
AnswerTimeout 8000
BaudRate 115200
DelayBetweenCharsMs 0
Handshaking Off
Parity None
StopBits 1

This adapter interfaces to the Yokogawa CSUX family of spinning disk confocal scanheads. There is no additional software needed. Connect the CSUX controller to an available serial port on your computer. You will need to define the Hub device, and assign the correct serial port to it.

The CSUX can have up to 2 filter wheels. If you have only one, use the default number (1) for your filter wheel. DriveSpeed has a property called 'AutoAdjust'. You can enter here the exposure time (in milliseconds) you are using. The scanhead will then calculate a disk speed at which you will get a integral number of disk rotations during that exposure. This can help reduce artefacts in your image. If you change the disk speed using other controls, the value for this property will go back to 0.

Note
After changing drive speed, the software will wait until the speed is within ~5% of its final target. Use the "Refresh" button to get the correct speed.

As of version 1.3.38, MIcro-Manager supports the bright field bypass option. There is no support for the second camera port option (although control can easily be added by anyone having access to such a scan-head).

If you experience a very uneven field of illumination in the CSUX, please check out CSU-X_fiber_rotation page.

The CSU-X1 requires a null modem cable.


© Micro-Manager : Vale Lab, UCSF 2006-2011 | All Rights Reserved | Contact