Description

Designed to handle standard SBS well plates from 24, 96, 384 and 1536 wells.

The plugin will calculate the correct coordinates for each well position. You can then iterate through each well, starting at the first well. You may either zero position of the XY stage on the microscope for the center of well A1 to coordinates to {0,0} or you can pass the X and Y offsets.

The plugin will move the stage down or up each column, until the last row, before moving to the next column and reversing direction. The zig-zag path chosen for moving the stage is the fastest path on an Zeiss AxioObserver Z1.

You can change the starting well, or the ending well to only work on a subregion.

Setup

big.jar must be installed and imported into the bsh script.

addClassPath("Big.jar");
import edu.umassmed.big.SBSPlate;

Class Functions

Most functions expect well positions in X, Y coordinates, where X is the long axis, and Y is the short axis of the well plate. X might have a range from 1 to 48 and Y might have a range from 1 to 32, depending upon plate size. Functions with Position in the name return coordinates in microns relative to well A1.

SBSPlate()
Initializes the plugin and uses 96 as the size of the well plate.
SBSPlate(int size)
Initializes the plugin for use with well plate of a set size. Acceptable sizes are 24, 96, 384, and 1536. Assumes Well A1 is at
coordinates 0,0.
SBSPlate(int size, double X, double Y)
This also sets the X and Y stage coordinates of well A1. Only use one initialization SBSPlate();
void setFirstWell(int x, int y)
Changes the starting position from {1,1} to {x,y}. Index is 1 based.
void setLastWell(int x, int y)
Changes the ending position from {1,1} to {x,y}. Index is 1 based.
double[] getWellPosition (int x, int y)
Returns the absolute position in microns relative to well A1. Calling X and Y is 1 indexed but returns X and Y 0 indexed position in microns.
String getWellLabel(int well)
Returns the SBS name of the well as a String formatted as letter and number. For example, A01. If you are only using a subregion, then you only need to pass the well number of that subregion, with well 1 being the start of the subregion. The function will calculate the correct SBS label of that subregion based on where that subregion is on the well plate.
int getNumberOfWells()
Returns the number of wells in the entire plate, or the number of wells in a subregion.
double[] getNextPosition()
Returns the {x,y} coordinates in microns of the next well. If at the last well, it will return the first well's coordinates.

Example Micro-Manager BSH Script

import edu.umassmed.big.SBSPlate;

gui.closeAllAcquisitions();
gui.clearMessageWindow();

xystage = mmc.getXYStageDevice();

double[] position = {0,0}; // 0 indexed in microns.
//
// Move stage to 0,0 if your stage is 0,0 at well A1
//
mmc.setXYPosition(xystage,position[0],position[1]);
mmc.waitForDevice(xystage);
//
// or get your current X,Y position of well A1
//
double startX = mmc.getXPosition(xystage);
double startY = mmc.getYPosition(xystage);

// Pass the well plate type, and optionally pass the X and Y offset of well A1
SBSPlate plate = new SBSPlate(96,startX, startY);

numberOfWells = plate.getNumberOfWells();

for (m = 1; m <= numberOfWells; m++) {
  // code to take an image could be here


  // move the stage to the next well
  position = plate.getNextPosition(); // 0 indexed in mcrons
  mmc.setXYPosition(xystage,position[0],position[1]);
  mmc.waitForDevice(xystage);
}
// return stage to starting position, either to {0,0} or {startX, startY}
mmc.setXYPosition(xystage,startX,startY);
mmc.waitForDevice(xystage);


--Karl Bellve, Biomedical Imaging Group, University of Massachusetts 20:02, 9 February 2011 (UTC)