== Description ==
 
== Description ==
  
pgFocus is an open source and open hardware focus stabilization device developed by Karl Bellvé at the [http://big.umassmed.edu Biomedical Imaging Group] (PMM, University of Massachusetts Medical School).
+
pgFocus is an open source and open hardware focus stabilization device developed by Karl Bellvé at the [http://big.umassmed.edu Biomedical Imaging Group] (Progam in Molecular Medicine, University of Massachusetts Medical School).
  
 
The pg in pgFocus is short for "Pretty Good". pgFocus isn't "Perfect", or "Definite" but it is pretty good!
 
The pg in pgFocus is short for "Pretty Good". pgFocus isn't "Perfect", or "Definite" but it is pretty good!
Line 10: Line 10:
  
 
pgFocus is connected to a computer via a USB port. USB is used to power and to communicate with pgFocus.
 
pgFocus is connected to a computer via a USB port. USB is used to power and to communicate with pgFocus.
 +
 +
==== Power ====
 +
 +
pgFocus uses the USB +5V power to generate +10V and -5V power. The actual output voltage for +10V power line is lower due to the forward voltage drop of the diodes required for the boost converter IC. However, one should never be running their piezo Z controller near their max voltage input range anyway. Just be aware that voltage signals near +10V sent through pgFocus may be reduced.
 +
 +
The -5V is higher (more positive) because of the same reason. Our piezo Z controller accepts -2V to +10V, so this isn't a significant issue on the -5V power line.
 +
 +
==== Digital to Analog Converter ====
 +
 +
pgFocus uses a ±5V 14bit DAC to modify incoming piezo Z signal. This is the upper and lower voltage limit that can be used to adjust the piezo Z signal. The DAC has a resolution of 0.6 mV. Based on our piezo Z controller, this works out to a 6 nM step size at the objective.
 +
 +
==== Analog to Digital Converter ====
 +
 +
pgFocus uses a ±10V ADC to sample the incoming piezo Z signal. This is to allow pgFocus to automatically adjust the focus set point, without user intervention. This feature enables pgFocus to have 3D focus correction.
 +
 +
The incoming piezo signal is sampled and multiplied by 2. This means that the ±10V ADC becomes effectively a ±5V ADC, matching the ±5V DAC.
 +
 +
However, this means the lower limit for sampling the incoming piezo Z signal is actually -2.5V, which becomes -5V, due to the gain of 2, at the ADC. This is acceptable because you should never drive your piezo Z controller below -2V. Therefore the effective ADC sampling range is actually -2.5V to +5V.
 +
 +
When pgFocus samples the incoming piezo Z signal, it will round off the results to the nearest 50nM step since most people move their objectives at 50, 100 or 250nm increments. This will allow the adjustment of the focus set point more accurately based on the distance from the original focus set point.
 +
 +
==== Linear Light Array====
 +
 +
Additionally, the Linear Light Array represents a physical limit. The return beam must fall within its 128 pixels. The length of the 128 pixels is about 8 mm. In our setup, the 8 mm length of the Linear Light Array is equivalent to about 25 to 30µM range at with a 60X objective. Longer Linear Light Arrays maybe used in future versions of pgFocus.
 +
 +
The Linear Light Array's location on the PCB is in the exact vertical center of the board. This allows you to put pgFocus in either vertical orientation using the same case.
 +
 +
==== Digital Expansion Port ====
 +
 +
pgFocus makes accessible out 8 digital I/O pins compatible with D0 to D7 pins on the Arduino Leonardo. These pins can be used for other light sensors, or perhaps a hardware switch to turn on/off focus, or to move the focus set point up or down.
 +
 +
+5V and ground lines are also made available. However, since USB is powering pgFocus, power maybe limited. Future versions of pgFocus might have a 5V DC jack if power is a limitation in the current design.
  
 
Eagle files are located at [https://github.com/kbellve/pgFocus Github].
 
Eagle files are located at [https://github.com/kbellve/pgFocus Github].
Line 15: Line 47:
 
'''Warning:''' PCB design has not been verified yet and may not be final.
 
'''Warning:''' PCB design has not been verified yet and may not be final.
  
=== Optics ===
+
==== Optics ====
  
 
pgFocus uses a laser beam that is reflected off the cover glass/water interface to monitor the microscope focus. Hence, an objective with an NA of at least 1.45 is required.
 
pgFocus uses a laser beam that is reflected off the cover glass/water interface to monitor the microscope focus. Hence, an objective with an NA of at least 1.45 is required.
Line 39: Line 71:
 
pgFocus software has not be released yet since it was designed for the prototype version of pgFocus rather than this version. It will be released as soon as possible.
 
pgFocus software has not be released yet since it was designed for the prototype version of pgFocus rather than this version. It will be released as soon as possible.
  
=== Calibration Mode ===
+
==== Calibration Mode ====
  
 
pgFocus needs to understand the relationship between the movement of the laser beam return signal seen on the Linear Light Array and the movement of the objective. This depends upon the location of pgFocus in relation to the objective. To determine this relationship, pgFocus has a calibration mode that moves the objective up and then down, while reading the position of the return laser beam. This information is plotted and a linear regression analysis is performed to determine the slope. The slope is the scale factor that pgFocus uses for calculations and for focus adjustments.
 
pgFocus needs to understand the relationship between the movement of the laser beam return signal seen on the Linear Light Array and the movement of the objective. This depends upon the location of pgFocus in relation to the objective. To determine this relationship, pgFocus has a calibration mode that moves the objective up and then down, while reading the position of the return laser beam. This information is plotted and a linear regression analysis is performed to determine the slope. The slope is the scale factor that pgFocus uses for calculations and for focus adjustments.
Line 71: Line 103:
 
As you can see in the bottom graph, focus position has changed even after the doors were closed due to the perturbations introduced into the microscope. The focus position may eventually correct itself once the temperature gradient returns to its original gradient.
 
As you can see in the bottom graph, focus position has changed even after the doors were closed due to the perturbations introduced into the microscope. The focus position may eventually correct itself once the temperature gradient returns to its original gradient.
  
These a severe perturbations but pgFocus is very good at compensating for slow focal drift over time.
+
These are severe perturbations but pgFocus is very good at automatically compensating for slow focal drift over time.
  
 
== Parts ==
 
== Parts ==

Revision as of 08:34, 21 February 2013

Description

pgFocus is an open source and open hardware focus stabilization device developed by Karl Bellvé at the Biomedical Imaging Group (Progam in Molecular Medicine, University of Massachusetts Medical School).

The pg in pgFocus is short for "Pretty Good". pgFocus isn't "Perfect", or "Definite" but it is pretty good!

Hardware

pgFocus uses the same Atmel ATMEGA 32U4 as seen in the Arduino Leonardo.

pgFocus is connected to a computer via a USB port. USB is used to power and to communicate with pgFocus.

Power

pgFocus uses the USB +5V power to generate +10V and -5V power. The actual output voltage for +10V power line is lower due to the forward voltage drop of the diodes required for the boost converter IC. However, one should never be running their piezo Z controller near their max voltage input range anyway. Just be aware that voltage signals near +10V sent through pgFocus may be reduced.

The -5V is higher (more positive) because of the same reason. Our piezo Z controller accepts -2V to +10V, so this isn't a significant issue on the -5V power line.

Digital to Analog Converter

pgFocus uses a ±5V 14bit DAC to modify incoming piezo Z signal. This is the upper and lower voltage limit that can be used to adjust the piezo Z signal. The DAC has a resolution of 0.6 mV. Based on our piezo Z controller, this works out to a 6 nM step size at the objective.

Analog to Digital Converter

pgFocus uses a ±10V ADC to sample the incoming piezo Z signal. This is to allow pgFocus to automatically adjust the focus set point, without user intervention. This feature enables pgFocus to have 3D focus correction.

The incoming piezo signal is sampled and multiplied by 2. This means that the ±10V ADC becomes effectively a ±5V ADC, matching the ±5V DAC.

However, this means the lower limit for sampling the incoming piezo Z signal is actually -2.5V, which becomes -5V, due to the gain of 2, at the ADC. This is acceptable because you should never drive your piezo Z controller below -2V. Therefore the effective ADC sampling range is actually -2.5V to +5V.

When pgFocus samples the incoming piezo Z signal, it will round off the results to the nearest 50nM step since most people move their objectives at 50, 100 or 250nm increments. This will allow the adjustment of the focus set point more accurately based on the distance from the original focus set point.

Linear Light Array

Additionally, the Linear Light Array represents a physical limit. The return beam must fall within its 128 pixels. The length of the 128 pixels is about 8 mm. In our setup, the 8 mm length of the Linear Light Array is equivalent to about 25 to 30µM range at with a 60X objective. Longer Linear Light Arrays maybe used in future versions of pgFocus.

The Linear Light Array's location on the PCB is in the exact vertical center of the board. This allows you to put pgFocus in either vertical orientation using the same case.

Digital Expansion Port

pgFocus makes accessible out 8 digital I/O pins compatible with D0 to D7 pins on the Arduino Leonardo. These pins can be used for other light sensors, or perhaps a hardware switch to turn on/off focus, or to move the focus set point up or down.

+5V and ground lines are also made available. However, since USB is powering pgFocus, power maybe limited. Future versions of pgFocus might have a 5V DC jack if power is a limitation in the current design.

Eagle files are located at Github.

Warning: PCB design has not been verified yet and may not be final.

Optics

pgFocus uses a laser beam that is reflected off the cover glass/water interface to monitor the microscope focus. Hence, an objective with an NA of at least 1.45 is required. The laser beam wavelength should be different from that being used for science (808nm is a good choice) and must be coaligned with the TIRF beam. This can be done using a mirror, beam splitter and a short pass filter. Care must be taken to ensure that the fluorescence filter cube and other microscope optics will allow the return of the pgFocus beam back along its input path and back to the pgFocus detector and that the microscope optics will NOT allow the laser light to pass into the user's eyes or to reach the camera. Once the position and angle of the pgFocus beam is coincident with that of the TIRF beam a strong return pgFocus beam will be seen that moves with changes in microscope focus. The pgFocus detector array can now be aligned and centered on the beam.

Software

After the Arduino Leonardo was released, pgFocus was redesigned to take advantage of the bootloader, as well as the pin layout for the Arduino Leonardo. If you can write an Arduino sketch, you can modify the source code to pgFocus!

pgFocus software reports the following:

  • Light profile as seen by the Linear Light Array.
    • Use this to align the physical position of pgFocus with respect to the return laser beam.
  • Current centroid location of the light profile as seen by the Linear Light Array.
  • Microscope stability based on the running Standard Deviation(converted to nM movement at the objective).
  • Digital to Analog converter voltage output need to maintain current focus position.
  • Calibration mode result.

pgFocus software can do the following:

  • Lock focus at the current position.
  • Move focus up by a specified amount (currently set to +50nM).
  • Move focus down by a specified amount (currently set to -50nM).

pgFocus software has not be released yet since it was designed for the prototype version of pgFocus rather than this version. It will be released as soon as possible.

Calibration Mode

pgFocus needs to understand the relationship between the movement of the laser beam return signal seen on the Linear Light Array and the movement of the objective. This depends upon the location of pgFocus in relation to the objective. To determine this relationship, pgFocus has a calibration mode that moves the objective up and then down, while reading the position of the return laser beam. This information is plotted and a linear regression analysis is performed to determine the slope. The slope is the scale factor that pgFocus uses for calculations and for focus adjustments.

A calibration must be performed after first installing pgFocus, or if the distance between pgFocus or the objective changes (i.e. pgFocus is moved).

The user just needs to make sure the the return laser beam is aligned and centered on the Linear Light Array. pgFocus will take care of the rest and will restore the resulting calculation in its EPROM.

Performance

pgFocus is designed to track and control focus within ±3nM at 30Hz under optimal conditions (tested with a high NA 60X Objective plus 1.6X optivar). This is dependent upon the location of pgFocus with respect to the objective.

  • Precision is proportional to the distance pgFocus is from the objective.
  • Range is inversely proportional to the distance pgFocus is from the objective

Movie of pgFocus in action. 200nM beads were TIRF imaged with uManager's burst mode with a 60X objective with a 1.6X optivar.

The microscope is enclosed in an environmental chamber. The doors to the chamber were opened at image 300 and closed at image 600. You will see a disturbance to the focus at both times, but focus is soon corrected within moments. The focus is disturbed due to the change in the temperature gradient across the microscope caused by the open doors.

Below is the graph of pgFocus performance from the above movie.

Figure: pgFocus in action. Doors to a microscope environmental chamber were opened at image 300 and then closed at image 600.


As you can see in the above graph, even after the doors were close, a constant voltage adjustment had to be applied to keep focus at its original position.

Below is the graph of what happens when pgFocus isn't controlling focus stabilization during the same procedure.

Figure: pgFocus has been turned off, but the doors were opened and closed to the microscope chamber. Only plotting the Linear Light Array since no voltage is being applied to correct focus.

As you can see in the bottom graph, focus position has changed even after the doors were closed due to the perturbations introduced into the microscope. The focus position may eventually correct itself once the temperature gradient returns to its original gradient.

These are severe perturbations but pgFocus is very good at automatically compensating for slow focal drift over time.

Parts

Preliminary parts list. Parts may change at any time.

Name Current Voltage Reference Notes
TSL1401CL 5mA 5V Mouser Octopart 0.018 to 100ms Exposure Time
ATMEGA32U4 7mA 5V ATMEL Newark
AD8221ARZ 1mA -12V to 12V Analog Devices PDF
AD5531BRUZ 2mA -12V to 12V Analog Devices PDF
TC7660 20mA 5V to 10V, -5V Microchip PDF Newark 10kHz. Use 10uF caps
TC7660H 5V to 10V, -5V Microchip PDF Newark 120kHz. Use 1uF caps
AD7894-10 4mA -10V to 10V Analog Devices PDF Newark Serial (not SPI or I2C), Using GAIN on 2 with op amp for -5 to +5V range
ADR03BRZ 1mA 2.5V reference Analog Devices PDFNewark 2.5V Reference
ADTL082 1.2mA Analog Devices PDF Unity Gain Stable
RJ45-8PTH NA NA
USB Type B NA NA Newark
PTC Resettable Fuse 500mA 15V Newark
Schottky Diode 30mA 30V Digi-key 2-SMD, Flat Lead
Schottky Diode 200mA 40V Newark SOD-323
18pF Capacitor 50V Newark 0603
100pF Capacitor 50V Newark 0603
1000pF Capacitor 50V Newark 0603
0.01uF Capacitor 50V Newark 0603
0.1uF Capacitor 50V Newark 0603
10uF Capacitor 16V Newark 1206, ESR: 0.2ohm
10uF Capacitor 16V Newark PDF 1206, ESR: 3ohm
Enclosure Hammond PDF Part Number: 1455NC1201


Thanks

pgFocus couldn't have been designed without the support from the Program in Molecular Medicine at Umass, and the following Open Software and Open Hardware communities: