Platforms: All platforms (uses USB or RS232 Port)
 
Platforms: All platforms (uses USB or RS232 Port)
 
Devices: Omicron “…xX”-laser series
 
Devices: Omicron “…xX”-laser series
Since Version:  1.48 (64bit compatible since version 1.4.12)
+
Since Version:  1.4.8 (64bit compatible since version 1.4.12)
  
 
Default serial port settings: AnswerTimeout         2.5 seconds
 
Default serial port settings: AnswerTimeout         2.5 seconds

Revision as of 10:53, 5 December 2013

Omicron Laser Controller

Summary:	Omicron Laser Controller
Author:	        Jan-Erik Herche/Ralf Schlotter
License:	LGPL
Platforms:	All platforms (uses USB or RS232 Port)
Devices:	Omicron “…xX”-laser series
Since Version:  1.4.8 (64bit compatible since version 1.4.12)

Default serial port settings:	AnswerTimeout	        2.5 seconds
	                        BaudRate	        57600 (RS232 Port) 
                                                        500000 (USB Port)
	                        DelayBetweenCharsMS	0
	                        Handshaking	        Off
	                        Parity	                None
	                        StopBits	        1

The Omicron Laser Controller controls Omicron LuxX®, PhoxX® and BrixX® series lasers as well as lasers integrated in LightHUB® compact laser combiners and SOLE® laser light engines via their serial interfaces (USB Port - all models, RS232 Port - for LuxX® and BrixX® series lasers only available with an optional break out cable). In the hardware configuration wizard you will need to add one Omicron Laser Controller object for every “…xX”-laser device you would like to control.

The following laser features and operating modes can be controlled and/or monitored through µ-manager device/property browser:

Read only properties:

  • Device type
  • Serial number
  • Specified power
  • Wavelength
  • Workinghours
  • Temperature baseplate
  • Temperature diode
  • Error notification

Read/Write properties:

  • Laser operation select (turns the laser emission ON or Off)
  • Laser operatingmode select (CW, analogue and/or digital modulation, standby)
  • Laser CW sub-operatingmode select (ACC-Auto Current Control or APC-Auto Power Control; only available with connected LuxX or BrixX laser module)
  • Laser power set-point select (set the laser power level in [mW] or [ %])
  • Reset
Note
Laser emission will be automatically shut off at µ-Manager start up. Laser emission can be turned back on as desired by the “Laser Operation Select” property.
Note
Laser properties show the values of the last polling, simply click the “Refresh” button of the device/property browser to get the actual values. Adhoc communication of the laser is automatically disabled at µ-Manager start up.
Note
Laser emission will automatically shut off when µ-manager is shut down.

Below you can see a screenshot of an Omicron “…xX”-Laser Series module controlled via the device/property browser: