top bar

Difference between revisions of "Hamamatsu"

(Hamamatsu DCAM adapter)
Line 15: Line 15:
 
Two versions of the adapter are included with Micro-Manager: the "Hamamatsu" adapter is an older version, written by the Micro-Manager team at UCSF, the "HamamatsuHam" adapter is newer and written by engineers at Hamamatsu.  Please do use the HamamatsuHam adapter unless there are reasons not to.  
 
Two versions of the adapter are included with Micro-Manager: the "Hamamatsu" adapter is an older version, written by the Micro-Manager team at UCSF, the "HamamatsuHam" adapter is newer and written by engineers at Hamamatsu.  Please do use the HamamatsuHam adapter unless there are reasons not to.  
  
In the case of ORCA flash4.0, go straight to HamamatsuHam in the configuration. For ORCA flash4.0, trying Hamamatsu adapter first will not only get the error message but also prevent the correct communication for trying HamamatsuHam afterwards. Rebooting the computer reset such a case. Also, if the computer has gone through a cycle of sleep, rebooting the computer and the camera is needed.
+
In the case of ORCA Flash 4.0, please read the page on [[High Speed Acquisition in Micro-Manager]]. For the ORCA Flash 4.0, trying Hamamatsu adapter first will not only get the error message but also prevent the correct communication for trying HamamatsuHam afterwards. Rebooting the computer reset such a case. Also, if the computer has gone through a cycle of sleep, rebooting the computer and the camera is needed.
  
 
If Micro-Manager reports: "No camera found" while configuring Micro-Manager with a Hamamatsu camera, make sure that it works with the software provided by Hamamatsu.  On Windows, you can launch it by starting the Hamamatsu installer, click the "Others" button, and select "ExCap".  Make sure that Excap finds your camera and that you can "Snap" an image in ExCap.  If ExCap does not find the camera, it is likely that the wrong driver is installed for your Firewire Hamamatsu camera.  Open the Windows Device Manager (Start->Control Panel->System->Hardware->Device Manager). Find the Hamamatsu camera (it can be under "Imaging Devices").  Right click and select "Properties".  Make sure that the driver is from Hamamatsu.  If it is not (most likely alternative is that the driver is from Microsoft), uninstall the driver, and re-install, making sure that the Hamamatsu driver is installed this time.
 
If Micro-Manager reports: "No camera found" while configuring Micro-Manager with a Hamamatsu camera, make sure that it works with the software provided by Hamamatsu.  On Windows, you can launch it by starting the Hamamatsu installer, click the "Others" button, and select "ExCap".  Make sure that Excap finds your camera and that you can "Snap" an image in ExCap.  If ExCap does not find the camera, it is likely that the wrong driver is installed for your Firewire Hamamatsu camera.  Open the Windows Device Manager (Start->Control Panel->System->Hardware->Device Manager). Find the Hamamatsu camera (it can be under "Imaging Devices").  Right click and select "Properties".  Make sure that the driver is from Hamamatsu.  If it is not (most likely alternative is that the driver is from Microsoft), uninstall the driver, and re-install, making sure that the Hamamatsu driver is installed this time.

Revision as of 11:47, 26 September 2013

Hamamatsu DCAM adapter

Summary:Interfaces with Hamamatsu cameras
Author:Nenad Amodaj, with contributions from Nico Stuurman (Mac adaptation). Hamamatsu version of the adapter by Pat Gregorio (Hamamatsu)
License:LGPL (UCSF version, however, we have been asked not to distribute the source) and closed source (Hamamatsu version)
Platforms:Windows and Mac OS X
Devices:Hamamatsu DCAM compatible cameras (only tested with Firewire cameras)
Example Config File:Media:MMConfig_ham.cfg

This adapter makes Micro-Manager work with DCAM-compatible Hamamatsu cameras. You will need to download and install a recent version of the DCAM driver.

Note
DCAM version dated March 2011 does not work with Micro-Manager. The August 2011 release works fine.

Two versions of the adapter are included with Micro-Manager: the "Hamamatsu" adapter is an older version, written by the Micro-Manager team at UCSF, the "HamamatsuHam" adapter is newer and written by engineers at Hamamatsu. Please do use the HamamatsuHam adapter unless there are reasons not to.

In the case of ORCA Flash 4.0, please read the page on High Speed Acquisition in Micro-Manager. For the ORCA Flash 4.0, trying Hamamatsu adapter first will not only get the error message but also prevent the correct communication for trying HamamatsuHam afterwards. Rebooting the computer reset such a case. Also, if the computer has gone through a cycle of sleep, rebooting the computer and the camera is needed.

If Micro-Manager reports: "No camera found" while configuring Micro-Manager with a Hamamatsu camera, make sure that it works with the software provided by Hamamatsu. On Windows, you can launch it by starting the Hamamatsu installer, click the "Others" button, and select "ExCap". Make sure that Excap finds your camera and that you can "Snap" an image in ExCap. If ExCap does not find the camera, it is likely that the wrong driver is installed for your Firewire Hamamatsu camera. Open the Windows Device Manager (Start->Control Panel->System->Hardware->Device Manager). Find the Hamamatsu camera (it can be under "Imaging Devices"). Right click and select "Properties". Make sure that the driver is from Hamamatsu. If it is not (most likely alternative is that the driver is from Microsoft), uninstall the driver, and re-install, making sure that the Hamamatsu driver is installed this time.

This adapter does not work with all cameras that are DCAM compatible. All firewire cameras that we tested (AG, ERG2, and R2) work alright under Windows, some models had problems on some Intel Apple Macs (share your successes!). The EM cameras (C9100-12 and C9100-13) work both on Windows and Mac (tested with MM version 1.2.31). Other cameras with the Phoenix card may also work. In general, try before you buy, and please provide feedback here about successes and failures.

The ORCA-D2 was tested by us December 2011. It works well with the HamamatsuHam adapter and the August 2011 DCAM driver. Images from both CCDs appear as a single image in Micro-Manager. Use the SplitView plugin to display the two CCD images as an overlay. In the future, it will be possible to automatically apply the SplitView transformation to images acquired using Multi-Dimensional Acquisition. If you do not want the images to be acquired simultaneously, use property "ACTIVE VIEW INDEX" of the adapter. A setting of "0" will give you both images, setting "1" will give you only the image from one CCD, and setting "2" will give you the image from the other CCD.

--Nico 12:38, 15 July 2008 (PDT), last updated --Nico 17:09, 26 August 2009 (UTC)

Click here to search for Hamamatsu on the Micro-Manager mailing list archive.

© Micro-Manager : Vale Lab, UCSF 2006-2011 | All Rights Reserved | Contact