top bar

Difference between revisions of "AndorLaserCombiner"

(Add direct links to drivers since the ALC_SDK is a whopping 350 MB, and is both download restricted and not regularly updated on MyAndor.)
(Cleanup installation, detail fix for Prolific driver hang and link USB suspent to new PC setup wiki page.)
Line 16: Line 16:
 
|- valign="top"
 
|- valign="top"
 
!Platforms:
 
!Platforms:
|Windows XP 32bit<br>Windows 7 32bit and 64bit<br>(No Mac or Linux driver available)
+
|Windows (No Mac or Linux driver available)
 
|- valign="top"
 
|- valign="top"
 
!Devices:
 
!Devices:
Line 28: Line 28:
  
 
== Installation ==
 
== Installation ==
µManager 1.4.16 onward bundles all required libraries.  For versions than 1.4.16_20131025 follow the [[#Library installation for MM versions older than 1.4.16]].
 
  
If you're setting up a brand new computer, continue on to [[#Configure ALC_SDK dependencies]]
+
In most cases the ALC will work "out of the box".
  
== Hardware Configuration ==
+
But if you're setting up a brand new computer, continue on to [[#Configure ALC_SDK dependencies]], and for Micro-Manager versions than nightly build 1.4.16_20131025 follow the [[#Library installation for MM 1.4.15 and earlier]].
 +
 
 +
== Hardware Configuration Wizard ==
  
 
Description of devices available in the Hardware Configuration Wizard:
 
Description of devices available in the Hardware Configuration Wizard:
Line 107: Line 108:
 
=== No laser, but laser emission LEDs are illuminated ===
 
=== No laser, but laser emission LEDs are illuminated ===
 
{| class="wikitable"
 
{| class="wikitable"
!Description
+
!style="width: 20%;"|Description
 
|The hardware appears to be operating correctly, but there is no laser output.  The camera is running live, feeding the Active Blanking signal.
 
|The hardware appears to be operating correctly, but there is no laser output.  The camera is running live, feeding the Active Blanking signal.
|-
+
|- valign="top"
 
!Cause
 
!Cause
 
|DEMO-BOARD device listed in InstaCal
 
|DEMO-BOARD device listed in InstaCal
|-
+
|- valign="top"
 
!Workaround
 
!Workaround
 
|Delete the DEMO-BOARD device by opening InstaCal from <code>Start Menu > Programs > Measurement Computing</code>
 
|Delete the DEMO-BOARD device by opening InstaCal from <code>Start Menu > Programs > Measurement Computing</code>
Line 119: Line 120:
 
=== Microsoft Visual C++ error(s) opening HCW ===
 
=== Microsoft Visual C++ error(s) opening HCW ===
 
{| class="wikitable"
 
{| class="wikitable"
!Description
+
!style="width: 20%;"|Description
 
|Opening the <code>Hardware Configuration Wizard</code> shows one or more of the following error windows, and closing the window immediately crashes µManager:
 
|Opening the <code>Hardware Configuration Wizard</code> shows one or more of the following error windows, and closing the window immediately crashes µManager:
 
[[File:Alc_mscv_error.jpg|frame|Figure: Error due to incorrect internal laser power setting]]
 
[[File:Alc_mscv_error.jpg|frame|Figure: Error due to incorrect internal laser power setting]]
|-
+
|- valign="top"
 
!Cause
 
!Cause
 
|Bug in ALC_SDK (Laser power setting exceeding rated value) <!-- TODO: Add Andor's Bugzilla bug number -->
 
|Bug in ALC_SDK (Laser power setting exceeding rated value) <!-- TODO: Add Andor's Bugzilla bug number -->
|-
+
|- valign="top"
 
!Workaround
 
!Workaround
 
|Request Andor support to remotely connect to each laser head and adjust the "set power" (using the <code>p=</code> serial command) to the rated value of the laser head.
 
|Request Andor support to remotely connect to each laser head and adjust the "set power" (using the <code>p=</code> serial command) to the rated value of the laser head.
 
|}
 
|}
  
=== Hang during acquisition ===
+
=== Hang due to USB power saving ===
 
{| class="wikitable"
 
{| class="wikitable"
!Description
+
!style="width: 20%;"|Description
|Micro-Manager hangs during acquisition
+
|Micro-Manager hangs during acquisition.
|-
+
|- valign="top"
!Cause 1
+
!Cause
|Windows USB power saving is enabled
+
|Windows Vista and newer enable USB power saving, even in the "High Performance" power plan, by default.
|-
+
|- valign="top"
!Workaround 1
+
!Fix
|Disable Windows USB power saving
+
|[[Personal computer setup#Power_Options|Disable Windows "USB selective suspend setting"]].
|-
+
|}
!Cause 2
+
 
|Buggy Prolific driver (version 3) bundled by Windows 7
+
=== Hang due to incorrect Prolific USB-Serial driver ===
|-
+
{| class="wikitable"
!Workaround 2
+
!style="width: 20%;"|Description
|Install Prolific driver  [[:File:PL2303_DriverInstaller_v1413_20110219.gz|version 2.0.13.130]] and delete newer versions from the Device Manager.
+
|Micro-Manager hangs during acquisition.
 +
|- valign="top"
 +
!Cause
 +
|Buggy Prolific driver (version 3) bundled by Windows 7.  In the Device Manager when you right-click on the Prolific serial port:
 +
  Ports (COM & LPT)
 +
        ...
 +
        Prolific USB-to-Serial Comm Port (COM201)
 +
        ...
 +
 
 +
... it will show a newer driver version than 2.0.13.130.
 +
|- valign="top"
 +
!Workaround
 +
|
 +
# Install Prolific driver  [[:File:PL2303_DriverInstaller_v1413_20110219.gz|version 2.0.13.130]]
 +
# Disable Microsoft server driver installation.  When Windows says it's searching for drivers, it fetches them from the Microsoft server.  Turn off this driver updating behavior because if e.g. the USB bus glitches for whatever reason it will try to reinstall the buggy version 3 drivers (you have to do this step even if you have Windows updates disabled) via the Start menu > (Type in the "Search programs and files" box) change dev > Change device installation settings > No, Let me choose what to do > Never install driver from Windows Update
 +
# In your Device Manager, right-click on the Prolific ports and check the driver version.  The good version is 2.0.13.130.  Most likely you will see something like 3.4.48.272.  Right-click on the Prolfic entry in the Device Manager > Uninstall > Delete the driver software for this device > OK.  Repeat for all Prolific COM20x ports.  Then to pick up the next available driver in Windows, in the Device Manager click Action > Scan for hardware changes).  Until you get 2.0.13.130, repeat the above of uninstall+deleting the driver.  You will then need to reassign the COM ports to 20x which is detailed in [[AndorLaserCombiner#Configure_ALC_SDK_dependencies]].
 +
{{Note|Andor's ALCs have since switched to using FTDI USB-serial convertors because Prolific considers the PL-2303 an obsolete product and Prolific won't fix the issue with Microsoft distributing faulty drivers.}}
 
|}
 
|}
  

Revision as of 17:24, 27 June 2014

Summary: Interfaces with Andor laser combiner
Author: Karl Hoover, UCSF - µManager device adapter
Maintainer: Pariksheet Nanda, Andor - this documentation and µManager device adapter
Contributors: Ed Simmons, ES Imaging - patch for GUI TTL outputs
Pedro Almada, documentation fixes
License: BSD
Platforms: Windows (No Mac or Linux driver available)
Devices: LC-401, LC-501, LC-601 (only 4 laser LC-601 systems supported; 5 and 6 line support coming soon)
Since version 1.3.47


Installation

In most cases the ALC will work "out of the box".

But if you're setting up a brand new computer, continue on to #Configure ALC_SDK dependencies, and for Micro-Manager versions than nightly build 1.4.16_20131025 follow the #Library installation for MM 1.4.15 and earlier.

Hardware Configuration Wizard

Description of devices available in the Hardware Configuration Wizard:

Available Devices Description
AndorLaserCombiner Main device required for your hardware configuration.
Piezostage Enable Andor's analog BNC control for your Piezo Z device. Generally, confocal systems have this configuration since it moves the Piezo faster than RS-232 serial control. If you are unsure about this, look at the front of your Andor Precision Control Unit (PCU)- if you see a BNC cable connected from A5 of the PCU to your Piezo Z controller, you should enable this option. If you see the A5 BNC sealed off, you have a 6 laser system that therefore does not have this feature.
Initialization Properties Description
MultiPortUnitPresent Leave as the default "0" (false) if you only see a single fiber optic connected horizontally to the front of the laser combiner. Otherwise select "1" (true) if you see the multi fiber optic switcher attachment.

For Auto-Shutter setting in Step 3 of 6, set the Default Shutter to the AndorLaserCombiner and in Step 4 of 6 use the default Delay [ms] of 0.0 since a fast AOTF gates the laser output.

Device Properties

Property Description
Enable Logical shutter to allow selection of laser line. It can also set the laser to TTL mode, if the laser supports it. TTL mode requires firmware 2. When creating laser imaging channels, you should add the Enable properties for each laser wavelength to turn on only the wavelength(s) you need.
SaveLifetime Allows lasers to be put in standby mode to preserve lifetime hours. Changes to this property are saved onto the device memory, so the setting is persistently between Micro-Manager sessions. Since going into and coming out of standby can take a few seconds, best practice is leave this property "global" by not including it in channels. Instead, one could add these properties to the System Startup preset, or just create shortcuts by creating groups with only the SaveLifetime setting(s).
IsLinear Reads whether linear correction algorithm is being applied to AOTF by PowerSetpoint, otherwise AOTF laser power output is sigmoid.
PowerSetpoint AOTF intensity setting. Actual power output may or may not be linear depending on the value of IsLinear.
PowerReadback Current laser head power output. Output of 0, could be due to laser being put in Standby using SaveLifetime, or a fault with the laser head. If power is more than a few percent lower than MaximumLaserPower, it also indicates a faulty laser head, but some lasers can take up to 5 minutes to warm up (most warm up in 2 minutes).
MaximumLaserPower Laser rated operating power in milli-Watts.
Hours Indicates laser expired life to plan warranty contracts. (Unfortunately this always shows a value of "0" because the feature is not yet implemented by the Andor ALC SDK)
WaveLength Wavelength of laser line.
LaserState Laser state: "0" indicates off, "1" warm up (if SaveLifetime is set to "PowerOn") or standby (if SaveLifetime is set to "Standby"), and "2" is stable output.
NLasers Number of lasers available.
LaserPort Multi fiber port unit control.
DOUT Digital 8-bit output from Precision Control Unit. The lowest 2 bits are not accessible if a multi fiber port unit is present.
DIN Digital 8-bit input from Precision Control Unit. Reads TTL input signal state.

Troubleshooting (known issues and fixes)

No laser, but laser emission LEDs are illuminated

Description The hardware appears to be operating correctly, but there is no laser output. The camera is running live, feeding the Active Blanking signal.
Cause DEMO-BOARD device listed in InstaCal
Workaround Delete the DEMO-BOARD device by opening InstaCal from Start Menu > Programs > Measurement Computing

Microsoft Visual C++ error(s) opening HCW

Description Opening the Hardware Configuration Wizard shows one or more of the following error windows, and closing the window immediately crashes µManager:
Figure: Error due to incorrect internal laser power setting
Cause Bug in ALC_SDK (Laser power setting exceeding rated value)
Workaround Request Andor support to remotely connect to each laser head and adjust the "set power" (using the p= serial command) to the rated value of the laser head.

Hang due to USB power saving

Description Micro-Manager hangs during acquisition.
Cause Windows Vista and newer enable USB power saving, even in the "High Performance" power plan, by default.
Fix Disable Windows "USB selective suspend setting".

Hang due to incorrect Prolific USB-Serial driver

Description Micro-Manager hangs during acquisition.
Cause Buggy Prolific driver (version 3) bundled by Windows 7. In the Device Manager when you right-click on the Prolific serial port:
  Ports (COM & LPT)
        ...
        Prolific USB-to-Serial Comm Port (COM201)
        ...

... it will show a newer driver version than 2.0.13.130.

Workaround
  1. Install Prolific driver version 2.0.13.130
  2. Disable Microsoft server driver installation. When Windows says it's searching for drivers, it fetches them from the Microsoft server. Turn off this driver updating behavior because if e.g. the USB bus glitches for whatever reason it will try to reinstall the buggy version 3 drivers (you have to do this step even if you have Windows updates disabled) via the Start menu > (Type in the "Search programs and files" box) change dev > Change device installation settings > No, Let me choose what to do > Never install driver from Windows Update
  3. In your Device Manager, right-click on the Prolific ports and check the driver version. The good version is 2.0.13.130. Most likely you will see something like 3.4.48.272. Right-click on the Prolfic entry in the Device Manager > Uninstall > Delete the driver software for this device > OK. Repeat for all Prolific COM20x ports. Then to pick up the next available driver in Windows, in the Device Manager click Action > Scan for hardware changes). Until you get 2.0.13.130, repeat the above of uninstall+deleting the driver. You will then need to reassign the COM ports to 20x which is detailed in AndorLaserCombiner#Configure_ALC_SDK_dependencies.
Note
Andor's ALCs have since switched to using FTDI USB-serial convertors because Prolific considers the PL-2303 an obsolete product and Prolific won't fix the issue with Microsoft distributing faulty drivers.

Resources

Laser Combiner questions

For contacting Andor product support regarding any questions, use this Andor Product Support link

Click here to search for Andor on the Micro-Manager mailing list archive.

Development support

This driver is not actively maintained by the Andor software team. For feature requests or bug reports, please liaison with µManager developers on the mailing list and Pariksheet Nanda.


Appendix

Library installation for MM 1.4.15 and earlier

  • In older versions of Micro-Manager, the ALC_SDK library files were not included in the main Micro-Manager directory. For convience the files have been uploaded (with permission) to this wiki:
ALC Driver Pack Visual C++ 2008 redist
(Only for MM 1.4.14 and earlier)
32-bit ALC_SDK Libraries 2.0 (32-bit) vcredist_x86.exe
64-bit ALC_SDK Libraries 2.0 (64-bit) vcredist_x64.exe
         C:\ProgramData\Andor_ALC_library\
  • If you already have control of your ALC in another software, your ALC_SDK installation is complete because all the Windows device drivers would be installed. Otherwise continue on to #Configure ALC_SDK dependencies.

Configure ALC_SDK dependencies

  • There are 3 dependency drivers:
    • DeVaSys (for front USB control to enable interlocks)
    • Prolific PL-2303 USB (for laser serial communication). On the ALC-601 these have been replaced with standard Tronix FTDI drivers which don't require driver installation as they are bundled with Windows 7.
    • Measurement Computing (for PCI DAC card)
Install them from:
OS version DeVaSys Prolific Measurement Computing
32-bit UsbI2cIo 5.00 (32-bit) Prolific 2.0.13.130 InstaCal 6.10
64-bit UsbI2cIo 5.00 (64-bit)
  • The devices will be detected in the Device Manager as:
     DAS Component
        PCIM-DDA06/16
     DeVaSys USB Interfaces
        USB-I2C/IO Board
     ...
     Ports (COM & LPT)
        ...
        Prolific USB-to-Serial Comm Port
        ...
  • Run Instacal from Windows Start Button > All Programs > Measurement Computing. You will see the "Plug and Play Board Detection" window finding your PCIM-DDA06/16 card. Click OK and Exit Instacal. This has now written your DAC card configuration file.
  • Open the Device Manager to check the COM port numbers for the USB-to-Serial COM ports built into the ALC. Andor's library reserves specific port numbers for the combiner and expects them to be COM201-206. Change the port numbers as follows:
    • Unplug the blue USB cable from the front of the ALC and check if any such ports remain, to confirm if you have other Prolific USB-to-Serial devices attached. If see some still left, make a note of their COM numbers so that you don't change them. Plug the ALC back in and note the new COM numbers.
    • For each of your reappeared COM ports, check the port physical location by (Right-click) > Properties. On Windows 7, the "Port" is immediately visible under Location in the General tab. Search the Properties of all the Prolific USB-to-Serial devices until you see Port_#0001. That port will need to be changed to COM 201. Note the corresponding Hub_# number of that port.
    • Reassign your COM ports number by (Right-click) > Properties > Port Settings (tab) > Advanced > change the COM port number to 201
    • With your Hub number of COM 201 noted above, reassign the remaining Port_#0002, Port_#0003, Port_#0004 for that Hub as COM 202, 203, 204 as above.
    • If your sled has 5 or 6 lasers you will be left with two more Prolific USB-to-Serial ports. You will need to reassign Port_#0002, Port_#0003 as COM 205 and 206, respectively
  • Restart the computer to apply the COM port changes